Dimensions W x H x D (with COSY and 3-Vaporizer Mount) approx. | 40.7 x 52 x 33 in (103.5 x 132 x 84 cm) |
Dimensions W x H x D (without COSY, with 2-Vaporizer Mount) approx. | 30 x 52 x 33 in (76 x 132 x 84 cm) |
Power supply (non-configurable) | 100–240 VAC, 50/60 Hz, 70 VA, 2.3 A max. |
Operation time with fully charged batteries | up to 120 minutes, minimum 45 minutes |
Ventilator E-Vent | Electronically controlled, electrically driven |
Operating modes | |
Standard | Manual/Spontaneous; Volume Control (VC) |
Option | Pressure Control (PC); Pressure Support (PS); Synchronized Volume Controlled Ventilation with PS (SIMV/PS) |
Breathing frequency | 4 to 60 bpm Max. |
Minute Volume (MV) | 99 L/min |
Positive end-expiratory pressure (PEEP) | 0–20 cmH2O |
Inspiration/expiration ratio | (Ti:Te) 4 : 1 to 1 : 4 |
Pressure limiting (PMAX) | 15–70 cmH2O |
Tidal Volume (VT) | 20–1400 mL in Volume Control; 20–1100 mL in SIMV/PS |
Inspiratory pause (TIP | 0–50 % |
SIMV inspiratory time (TINSP) | 0.3–4.0 sec |
Inspiratory pressure (PINSP) | PEEP +5 to 65 cmH2O |
Inspiratory Flow (InspFlow) | 10–75 L/min in Volume and Pressure Control; 10–85 L/min in |
Pressure Support Pressure Support Level (PPS) | PEEP +3 to 20 cmH2O |
Min. frequency for apnea-ventilation (Freq. Min.) | 3–20 bpm and “OFF” |
Trigger | 2–15 L/min |
Integrated safety functions | Sensitive Oxygen Ratio Controller (S-ORC) provides a minimum O2 concentration of 23 % in an O2/N2O mixture. N2O cut-off if O2 fresh gas valve is closed or if O2 flow is less than 0.2 L/min. Audible and visual (flashing red LED) indication in case O2 pressure drops below 20 psi (1.38 bar) ± 4 psi (0.27 bar). In case of electricity and battery failure, manual ventilation, gas delivery and agent delivery are possible. Positive pressure relief valve opens at 75 ±5 cmH2O. Negative pressure relief valve opens at –7.5 to –9 cmH2O. |
Range of fresh gas flow indicators | 0.00 to 12.0 L/min |
Total fresh gas flow meter | 0 to 10 L/min, calibrated with a mixture of 50 % O2 and 50 % N20 mixture |
O2 flush | at 87 psi (6 bar): max. 75 L/min; at 41 psi (2.8 bar): min 25 L/min |
Vaporizer mount | 2 and 3 position Auto Exclusion mount/2 position Selectatec* Optional off-line parking position for 2 vaporizers |
Control screen | 6.5” (16.5cm) full color TFT display (option) |
Monitoring | Continuos monitoring of inspiratory O2 concentration, breathing frequency, tidal volume, minute volume, mean or plateau pressure, peak airway pressure, PEEP All fresh gas flow information is displayed as virtual flow tubes. |
Serial interface | 1 RS 232 (standard); 2nd RS 232 (option) |
Protocols | Vitalink and Medibus |
Data available for export | All fresh gas flow, ventilation and O2 data, flow curves and pressure curves |
Volume of CO2 absorber | Standard: 1.5 L reusable canister, Option: 1.2 L prefilled DrägerSorb CLIC absorber |
Volume of entire compact breathing system | 1.7 Liter + bag |
Additional accessories | Anesthetic Gas Scavenging, Endotracheal suction, auxiliary flow meter, low-flow kit with breathing system warmer, power outlet strip, pivotable shelf |